Calibration Capabilities |
|
|
|||
| MEASUREMENT CAPABILITIES | |||
| DIRECT VOLTS / SOURCE | |||
| TRACEABILITY | RANGE | RESOLUTION | UNCERTAINTY |
| +/- (ppm + uV) | |||
| Fluke 732A Ref. Std. | 1V | NA | 1.5 * |
| 1.018V | NA | 1.5 | |
| 10V | NA | 0.5 | |
| Fluke 5700A | 220mV | 10 nV | 8 + .75 ** |
| 2.2V | 100 nV | 7 + 1.2 | |
| 11V | 1 uV | 7 + 4 | |
| 22V | 1 uV | 7 + 8 | |
| 220V | 10 uV | 8 + 100 | |
| 1100V | 100 uV | 10 + 600 | |
| Fluke 5440A/B | 0V to 220mV | 0.01 uV | 6.5 + 0.5 ** |
| 0.22V to 2.2V | 0.1 uV | 6 + 1.0 | |
| 0V to 11V | 1uV | 2.5 + 5 | |
| 11V to 22V | 1uV | 2.5 + 8 | |
| 22V to 275V | 10uV | 4.5 + 100 | |
| 275V to 1100V | 100uV | 4.5 + 400 | |
| DIRECT VOLTS / MEASUREMENT | |||
| TRACEABILITY | RANGE | RESOLUTION | UNCERTAINTY |
| (ppm of reading)+(ppm of range) | |||
| HP3458A | 100mV | 10nV | 9 +3 *** |
| 1V | 10nV | 8 + 0.3 | |
| 10V | 100nV | 8 + 0.05 | |
| 100V | 1uV | 10 + 0.3 | |
| 1000 | 10 uV | 10 + 0.1 | |
| DIRECT AMPS / SOURCE | |||
| TRACEABILITY | RANGE | RESOLUTION | UNCERTAINTY |
| +/- ( ppm output + nA ) | |||
| 5700A | 220uA | 0.1nA | 55 + 10 ** |
| 2.2mA | 1nA | 55 + 10 | |
| 22mA | 10nA | 55 + 100 | |
| +/- (ppm output + uA ) | |||
| 220mA | 0.1uA | 65 + 1 | |
| 2.2A | 1A | 90 + 30 | |
| 5725A | 11A | 10nA | 350 + 480 |
| * 30 days | ** 6 months | *** 12 month | 1 |
| MEASUREMENT CAPABILITIES | |||
| DIRECT AMPS / MEASUREMENT | |||
| TRACEABILITY | RANGE | RESOLUTION | UNCERTAINTY |
| ( ppm reading + ppm range) | |||
| HP3458A | 100nA | 1pA | 30 + 400 *** |
| 1uA | 1pA | 20 + 40 | |
| 10uA | 1pA | 20 + 10 | |
| 100uA | 10pA | 20 + 8 | |
| 1mA | 100pA | 20 + 5 | |
| 10mA | 1nA | 20 + 5 | |
| 100mA | 10nA | 35 + 5 | |
| 1A | 100nA | 110 + 10 | |
| +/- ppm | |||
| Shunt ID# F202 | 5A | 0.25uA | 400 *** |
| Shunt ID# F203 | 10A | 0.5uA | 400 |
| Shunt ID# F204 | 25A | 2.5uA | 400 |
| Shunt ID# F237 | 100A | 5uA | 400 |
| Shunt ID# F057 | 750A | 75uA | 400 |
| Shunt ID# F058 | 1500A | 150uA | 400 |
| DC RATIO | |||
| TRACEABILITY | RANGE | ACCURACY | |
| Fluke 845AB | 1uV through 1000V | +/- ( 2% end-scale + 0.1uV ) | |
| MEASUREMENT CAPABILITIES | ||||
| ALTERNATING VOLTAGE / SOURCE | ||||
| TRACEABILITY | RANGE | RESOLUTION | FREQUENCY | UNCERTAINTY |
| Hz | +/- ( ppm output + uV ) | |||
| Fluke 5700A | 10 - 20 | 600 + 5 | ||
| 20 - 40 | 240 + 5 | |||
| 40 - 20k | 120 + 5 | |||
| 2.2mV | 1nV | 20k - 50k | 410 + 5 | |
| 50k - 100k | 950 + 8 | |||
| 100k - 300k | .13% + 15 | |||
| 300k - 500k | .18% + 30 | |||
| 500k - 1M | .36% + 30 | |||
| 10 - 20 | 600 + 6 | |||
| 20 - 40 | 240 + 6 | |||
| 40 - 20k | 120 + 6 | |||
| 22mV | 10nV | 20k - 50k | 410 + 6 | |
| 50k - 100k | 950 + 8 | |||
| 100k - 300k | .13% + 15 | |||
| 300k - 500k | .18% + 30 | |||
| 500k - 1M | .36% + 30 | |||
| 10 - 20 | 600 + 16 | |||
| 20 - 40 | 240 + 10 | |||
| 40 - 20k | 110 + 10 | |||
| 220mV | 100nV | 20k - 50k | 360 + 10 | |
| 50k - 100k | 900 + 30 | |||
| 100k - 300k | .11% + 30 | |||
| 300k - 500k | .18% + 40 | |||
| 500k - 1M | .36% + 100 | |||
| 10 - 20 | 600 + 100 | |||
| 20 - 40 | 180 + 30 | |||
| 40 - 20k | 85 + 7 | |||
| 2.2V | 1uV | 20k - 50k | 140 + 20 | |
| 50k - 100k | 280 + 80 | |||
| 100k - 300k | 480 + 150 | |||
| 300k - 500k | .12% + 400 | |||
| 500k - 1M | .24% + 1mV | |||
| 10 - 20 | 600 + 1mV | |||
| 20 - 40 | 180 + 300 | |||
| 40 - 20k | 85 + 70 | |||
| 22V | 10uV | 20k - 50k | 140 + 200 | |
| 50k - 100k | 280 + 400 | |||
| 100k - 300k | 600 + 1.7mV | |||
| 300k - 500k | .14% + 5mV | |||
| 500k - 1M | .30% + 9mV | |||
| Above uncertainty tables based on 12 month period. | 3 | |||
| MEASUREMENT CAPABILITIES | ||||
| ALTERNATING VOLTAGE /SOURCE ( Cont. ) | ||||
| TRACEABILITY | RANGE | RESOLUTION | FREQUENCY | UNCERTAINTY |
| Hz | +/- ( ppm output + mV ) | |||
| Fluke 5700A | 10 - 20 | 600 + 10 | ||
| 20 - 40 | 180 + 3 | |||
| 40 - 20k | 90 + 1 | |||
| 220V | 100uV | 20k - 50k | 250 + 4 | |
| 50k - 100k | 600 + 10 | |||
| 100k - 300k | .16% + 110 | |||
| 300k - 500k | .54% + 110 | |||
| 500k - 1M | 1.30% + 220 | |||
| 1100V | 1mV | 50 - 1k | 90 + 4 | |
| Fluke 5725A | 750V | 1mV | 30k - 50k | 600 + 11 |
| 50k - 100k | .23% + 45 | |||
| 40 - 1k | 90 + 4 | |||
| 1100V | 1mV | 1k - 20k | 165 + 6 | |
| 20k - 30k | 600 + 11 | |||
| ALTERNATING VOLTAGE / MEASUREMENT | ||||
| TRACEABILITY | RANGE | RESOLUTION | FREQUENCY | UNCERTAINTY |
| Hz | +/- (ppm of reading + uV ) | |||
| Fluke 5790A | 10 - 20 | 1700 + 1.3 | ||
| 20 - 40 | 740 + 1.3 | |||
| 40 - 20k | 420 + 1.3 | |||
| 2.2mV | .1uV | 20k - 50k | 810 + 2.0 | |
| 50k - 100k | 1200 + 2.5 | |||
| 100k - 300k | 2300 + 4.0 | |||
| 300k - 500k | 2400 + 8.0 | |||
| 500k - 1M | 3500 + 8.0 | |||
| 10 - 20 | 850 + 1.3 | |||
| 20 - 40 | 370 + 1.3 | |||
| 40 - 20k | 210 + 1.3 | |||
| 7mV | .1uV | 20k - 50k | 400 + 2.0 | |
| 50k - 100k | 600 + 2.5 | |||
| 100k - 300k | 1200 + 4.0 | |||
| 300k - 500k | 1300 + 8.0 | |||
| 500k - 1M | 2300 + 8.0 | |||
| 10 - 20 | 290 + 1.3 | |||
| 20 - 40 | 180 + 1.3 | |||
| 40 - 20k | 110 + 1.3 | |||
| 22mV | .1uV | 20k - 50k | 210 + 2.0 | |
| 50k - 100k | 310 + 2.5 | |||
| 100k - 300k | 810 + 4.0 | |||
| 300k - 500k | 860 + 6.0 | |||
| 500k - 1M | 1400 + 6.0 | |||
| Above uncertainty tables based on 12 month period. | 4 | |||
| MEASUREMENT CAPABILITIES | ||||
| ALTERNATING VOLTAGE / MEASUREMENT ( Cont. ) | ||||
| TRACEABILITY | RANGE | RESOLUTION | FREQUENCY | UNCERTAINTY |
| Hz | +/- ( ppm of reading + uV) | |||
| Fluke 5790A | 10 - 20 | 240 + 1.5 | ||
| 20 - 40 | 120 + 1.5 | |||
| 40 20k | 65 + 1.5 | |||
| 70mV | 1.0uV | 20k - 50k | 130 + 2.0 | |
| 50k - 100k | 260 + 2.5 | |||
| 100k - 300k | 510 + 4.0 | |||
| 300k - 500k | 670 + 8.0 | |||
| 500k - 1M | 1100 + 8.0 | |||
| 10 - 20 | 210 + 1.5 | |||
| 20 - 40 | 85 + 1.5 | |||
| 40 20k | 38 + 1.5 | |||
| 220mV | 1.0uV | 20k - 50k | 69 + 2.0 | |
| 50k - 100k | 160 + 2.5 | |||
| 100k - 300k | 250 + 4.0 | |||
| 300k - 500k | 380 + 8.0 | |||
| 500k - 1M | 1000 + 8.0 | |||
| 10 - 20 | 210 + 1.5 | |||
| 20 - 40 | 76 + 1.5 | |||
| 40 20k | 33 + 1.5 | |||
| 700mV | 10.0uV | 20k - 50k | 51 + 2.0 | |
| 50k - 100k | 79 + 2.5 | |||
| 100k - 300k | 180 + 4.0 | |||
| 300k - 500k | 300 + 8.0 | |||
| 500k - 1M | 960 + 8.0 | |||
| +/- (ppm of reading ) | ||||
| 10 - 20 | 200 | |||
| 20 - 40 | 66 | |||
| 40 20k | 24 | |||
| 2.2V | 10.0uV | 20k - 50k | 46 | |
| 50k - 100k | 71 | |||
| 100k - 300k | 160 | |||
| 300k - 500k | 260 | |||
| 500k - 1M | 900 | |||
| 10 - 20 | 200 | |||
| 20 - 40 | 67 | |||
| 40 20k | 24 | |||
| 7V | 100.0uV | 20k - 50k | 48 | |
| 50k - 100k | 81 | |||
| 100k - 300k | 190 | |||
| 300k - 500k | 400 | |||
| 500k - 1M | 1200 | |||
| Above uncertainty tables base on 12 month period. | 5 | |||
| MEASUREMENT CAPABILITIES | ||||
| ALTERNATING VOLTAGE /MEASUREMENT ( Cont. ) | ||||
| TRACEABILITY | RANGE | RESOLUTION | FREQUENCY | UNCERTAINTY |
| Hz | +/- ( ppm of reading ) | |||
| Fluke 5790A | 10 - 20 | 200 | ||
| 20 - 40 | 67 | |||
| 40 - 20k | 27 | |||
| 22V | 1.0uV | 20k - 50k | 48 | |
| 50k - 100k | 81 | |||
| 100k - 300k | 190 | |||
| 300k - 500k | 400 | |||
| 500k - 1M | 1200 | |||
| 10 - 20 | 200 | |||
| 20 - 40 | 68 | |||
| 40 - 20k | 32 | |||
| 70V | 1.0uV | 20k - 50k | 57 | |
| 50k - 100k | 94 | |||
| 100k - 300k | 200 | |||
| 300k - 500k | 410 | |||
| 500k - 1M | 1200 | |||
| 10 - 20 | 200 | |||
| 20 - 40 | 68 | |||
| 40 - 20k | 31 | |||
| 220V | 10.0uV | 20k - 50k | 69 | |
| 50k - 100k | 98 | |||
| 100k - 300k | 210 | |||
| 300k - 500k | 500 | |||
| 10 - 20 | 200 | |||
| 20 - 40 | 99 | |||
| 700V | 10.0uV | 40 - 20k | 41 | |
| 20k - 50k | 130 | |||
| 50k - 100k | 500 | |||
| 10 - 20 | 200 | |||
| 1000V | 100.0uV | 20 - 40 | 99 | |
| 40 - 20k | 38 | |||
| 20k - 50k | 130 | |||
| Above uncertainty tables based on 12 month period | 6 | |||
| MEASUREMENT CAPABILITIES | ||||
| ALTERNATING CURRENT / SOURCE | ||||
| TRACEABILITY | RANGE | RESOLUTION | FREQUENCY | UNCERTAINTY |
| Hz | +/- (ppm output + nA ) | |||
| Fluke 5700A | 10 - 20 | 800 + 30 | ||
| 20 - 40 | 420 + 25 | |||
| 220uA | 1nA | 40 - 1k | 160 + 20 | |
| 1k - 5k | 700 + 1uA | |||
| 5k - 10k | .18% + 4uA | |||
| 10 - 20 | 800 + 50 | |||
| 20 - 40 | 420 + 40 | |||
| 2.2mA | 10nA | 40 - 1k | 160 + 40 | |
| 1k - 5k | 700 + 10uA | |||
| 5k - 10k | .18% + 4uA | |||
| 10 - 20 | 800 + 500 | |||
| 20 - 40 | 420 + 400 | |||
| 22mA | 100nA | 40 - 1k | 160 + 400 | |
| 1k - 5k | 700 + 10uA | |||
| 5k - 10k | .18% + 20uA | |||
| Hz | +/- ( ppm output + uA ) | |||
| 10 - 20 | 800 + 5 | |||
| 20 - 40 | 420 + 4 | |||
| 220mA | 1uA | 40 - 1k | 180 + 4 | |
| 1k - 5k | 700 + 50 | |||
| 5k - 10k | .18% + 100 | |||
| 40 - 1k | 750 + 40 | |||
| 2.2A | 10uA | 1k - 5k | 850 + 100 | |
| 5k - 10k | 1.0% + 200 | |||
| Fluke 5725 | 40 - 1k | 460 + 170 | ||
| 11A | 100uA | 1k - 5k | 950 + 380 | |
| 5k - 10k | .36% + 750 | |||
| ALTERNATING CURRENT /MEASUREMENT | ||||
| TRACEABILITY | RANGE | RESOLUTION | FREQUENCY | UNCERTAINTY |
| Hz | ( % reading + % range ) | |||
| HP3458A | 10 - 20 | 0.4 + 0.02 | ||
| 100uA | 100pA | 20 - 45 | 0.15 + 0.02 | |
| 45 - 100 | 0.06 + 0.02 | |||
| 100 - 1k | 0.06 + 0.02 | |||
| 10 - 20 | 0.4 + 0.02 | |||
| 20 - 45 | 0.15 + 0.02 | |||
| 45 - 100 | 0.06 + 0.02 | |||
| 1mA | 1nA | 100 - 5k | 0.03 + 0.02 | |
| 5k - 20k | 0.06 + 0.02 | |||
| 20k - 50k | 0.4 + 0.04 | |||
| 50k - 100k | 0.55 + 0.15 | |||
| Above uncertainty tables based on 12 month period. | 7 | |||
| MEASUREMENT CAPABILITIES | ||
| ALTERNATING CURRENT / MEASUREMENT ( Cont. ) | ||
| TRACEABILITY | RANGE | RESOLUTION |
| HP3458A | ||
| 10mA | 10nA | |
| 100mA | 100nA | |
| 1A | 1uA | |
| Y5021 | 2 - 20A | |
| RESISTANCE /SOURCE | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| Ohms | ppm | |
| 742A-1 | 1.0 | 8.0 |
| 742A-1.9 | 1.9 | 8.0 |
| 742A-10 | 10.0 | 8.0 |
| 742A-100 | 100.0 | 6.0 |
| 742A-1k | 1.0k | 6.0 |
| 742A-10k | 10.0k | 4.0 |
| 742A-19k | 19.0k | 4.0 |
| 742A-100k | 100.0k | 6.0 |
| 172A-1M | 1.0M | 8.0 |
| 742A-10M | 10.0M | 9.0 |
| 742A-19M | 19.0M | 10.0 |
| MEASUREMENT CAPABILITIES | ||
| RESISTANCE / MEASUREMENT | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| Ohm | +/- ppm | |
| Fluke 5450A | 1 | 110 |
| 1.9 | 100 | |
| 10 | 33 | |
| 19 | 31 | |
| 100 | 16 | |
| 190 | 15.5 | |
| 1k | 13.5 | |
| 1.9k | 13 | |
| 10k | 13 | |
| 19k | 12.5 | |
| 100k | 14 | |
| 190k | 13.5 | |
| 1M | 19 | |
| 1.9M | 19 | |
| 10M | 50 | |
| 19M | 56 | |
| 100M | 200 | |
| CAPACITANCE / SOURCE | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| uF | ( +/- )% | |
| Gen Rad 1409-X | 0.5 | 0.01 |
| Gen Rad 1409-T | 0.1 | 0.01 |
| Gen Rad 1409-L | 0.01 | 0.01 |
| Gen Rad 1409-F | 0.001 | 0.01 |
| INDUCTANCE / SOURCE | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| ( +/- ) % | ||
| Gen Rad 1482L | 99 mH | 0.03 |
| FREQUENCY / SOURCE | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| WWVB | ||
| Spectracom 8161 | DC - 2.1GHz | 5 x 10 |
| Fluke 6062 | ||
| MEASUREMENT CAPABILITIES | ||
| FREQUENCY / MEASUREMENT | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| HP8902A | 10 Hz to 2.9GHz | 1 x 10 |
| HP70000 System | ||
| PM6685R | 10 Hz to 300MHz | 5 x 10 |
| FLW SERVICE CORP. | ||
| MEASUREMENT CAPABILITIES | ||
| PRESSURE /VACUUM | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| ( +/- ) | ||
| CEC 6-201 | 1 - 5 psi | 0.01% Indicated |
| 100-500 psi | 0.02% Indicated | |
| AMETEK T-150-1 | 5 - 15,000 psi | 0.025% Indicated |
| AMETEK PK654 | 4 - 659" H2O | 0.015% Indicated |
| MERIAM 34FB2 | 0 - 20" H2O | 0.005% Indicated |
| HEISE GAUGE | 0 - 30" Hg | 0.05% |
| TEMPERATURE | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| F172 - S Thermocouple | 32 to 2000 Deg. F | <0.5 Deg. F |
| 2000 to 2650 Deg. F | <3.0 Deg. F | |
| Kaye IRTD-400 | -183 to 420 Deg. C | +/- 0.025 Deg. C |
| Kaye IRTD-500 | -183 to 500 Deg. C | +/- 0.01 Deg. C |
| Lindberg Temp. Chamber | Ambient to 1100 Deg. C | +/-1.0 Deg. C over 2" zone |
| Mod. 1100 | ||
| Kaye LTR-50 | -30 to 80 Deg. C | Well uniformity: 0.05 Deg. C |
| Kaye HTR-400 | 50 to 400 Deg. C | Well uniformity: 0.05 Deg. C |
| Kaye HTR-600 | 50 to 600 Deg. C | Well uniformity: 0.05 Deg. C |
| DEW POINT / HUMIDITY | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| General Eastern 1100DP * | -40 to 80 Deg. C | +/-0.3 Deg. C |
| * Used with a Blue M VP100 Relative Humidity Chamber | ||
| MEASUREMENT CAPABILITIES | ||
| DIMENSIONAL | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| Starrett / Webber 84 Pcs. | 0.05" to 4.0" | Grade A+ |
| MASS | ||
| TRACEABILITY | RANGE | UNCERTAINTY |
| F087 Troemner | 1mg to 100g | Class S |
| F10200 | 1g to 2Kg | Class C |
| F064 | 0.1lb to 38.0lb | Class C |
| F144 | 1lb to 1,000lb | Class F |
| TRACEABILITY | RANGE | UNCERTAINTY |
| ( +/- ) % | ||
| Waters Torque Tester | 0 to 40 in. ozs. | 0.5 |
| Bytrex Torque Analyzer | 0 to 1200 in. lbs | 1.0 |
| Torque Controls | 0 to 1000 ft. lbs | 1.0 |
| Richmont Torque Analyzer | 0 to 7500 ft. lbs | 2.0 |